Apogee Alta Test Report: Difference between revisions

From Telescope
Jump to navigationJump to search
Cderoo1 (talk | contribs)
Created page with '<table> <tr><td width="100"><b>Initial Temp</b></td><td width="100"><b>End Temp</b></td><td width="100"><b>Time</b></td></tr> <tr><td>22.7</td><td>0</td><td>49</td></tr> <tr><td>...'
 
Cderoo1 (talk | contribs)
No edit summary
Line 1: Line 1:
<table>
<table>
<tr><td width="100"><b>Initial Temp</b></td><td width="100"><b>End Temp</b></td><td width="100"><b>Time</b></td></tr>
<tr><td width="100"><b>Initial Temp</b></td><td width="100"><b>End Temp</b></td><td width="100"><b>Time (min)</b></td></tr>
<tr><td>22.7</td><td>0</td><td>49</td></tr>
<tr><td>22.7</td><td>0</td><td>49</td></tr>
<tr><td>0</td><td>-5</td><td>36</td></tr>
<tr><td>0</td><td>-5</td><td>36</td></tr>
Line 7: Line 7:
<br>
<br>


<tr><td width="100"><b>Temperature</b></td><td width="100"><b>Dark Current</b></td><td width="100"><b>Read Noise</b></td></tr>
<table>
<tr><td>0</td><td> </td><td> </td></tr>
<tr><td width="100"><b>Temperature</b></td><td width="100"><b>Dark Current </b></td><td width="100"><b>Read Noise</b></td></tr>
<tr><td> -5</td><td> </td><td> </td></tr>
<tr><td>0</td><td>0.078</td><td>6.935</td></tr>
<tr><td>-10</td><td> </td><td> </td></tr>
<tr><td> -5</td><td>0.041</td><td>6.706</td></tr>
<tr><td>-10</td><td>0.035</td><td>6.631</td></tr>
</table>
</table>
<br>
<br>
The read noise is the standard deviation of the master bias image (9 combined images). The dark current is the mean of the bias-subtracted, median-combined darks times the gain (1.3) divided by the integration time (300 seconds). The dark current is in electrons per pixel per second, and the read noise is in ADU r.m.s. All of these measurements are below the values given in the camera test report.

Revision as of 20:40, 31 July 2009

Initial TempEnd TempTime (min)
22.7049
0-536
-5-1039


TemperatureDark Current Read Noise
00.0786.935
-50.0416.706
-100.0356.631


The read noise is the standard deviation of the master bias image (9 combined images). The dark current is the mean of the bias-subtracted, median-combined darks times the gain (1.3) divided by the integration time (300 seconds). The dark current is in electrons per pixel per second, and the read noise is in ADU r.m.s. All of these measurements are below the values given in the camera test report.