Apogee Alta Test Report: Difference between revisions
From Telescope
Jump to navigationJump to search
No edit summary |
No edit summary |
||
| Line 1: | Line 1: | ||
<table> | <table> | ||
<tr><td width="100"><b>Initial Temp</b></td><td width="100"><b>End Temp</b></td><td width="100"><b>Time (min)</b></td></tr> | <tr><td width="100"><b>Initial Temp</b></td><td width="100"><b>End Temp</b></td><td width="100"><b>Time to cool (min)</b></td></tr> | ||
<tr><td>22.7</td><td>0</td><td>49</td></tr> | <tr><td>22.7</td><td>0</td><td>49</td></tr> | ||
<tr><td>0</td><td>-5</td><td>36</td></tr> | <tr><td>0</td><td>-5</td><td>36</td></tr> | ||
| Line 15: | Line 15: | ||
<br> | <br> | ||
All the temperatures are in Celsius. | |||
The read noise is the standard deviation of the master bias image (9 combined images). The dark current is the mean of the bias-subtracted, median-combined darks times the gain (1.3) divided by the integration time (300 seconds). The dark current is in electrons per pixel per second, and the read noise is in ADU r.m.s. All of these measurements are below the values given in the camera test report. | The read noise is the standard deviation of the master bias image (9 combined images). The dark current is the mean of the bias-subtracted, median-combined darks times the gain (1.3) divided by the integration time (300 seconds). The dark current is in electrons per pixel per second, and the read noise is in ADU r.m.s. All of these measurements are below the values given in the camera test report. | ||
Revision as of 13:21, 3 August 2009
| Initial Temp | End Temp | Time to cool (min) |
| 22.7 | 0 | 49 |
| 0 | -5 | 36 |
| -5 | -10 | 39 |
| Temperature | Dark Current | Read Noise |
| 0 | 0.078 | 6.935 |
| -5 | 0.041 | 6.706 |
| -10 | 0.035 | 6.631 |
All the temperatures are in Celsius. The read noise is the standard deviation of the master bias image (9 combined images). The dark current is the mean of the bias-subtracted, median-combined darks times the gain (1.3) divided by the integration time (300 seconds). The dark current is in electrons per pixel per second, and the read noise is in ADU r.m.s. All of these measurements are below the values given in the camera test report.